Development of a rapid profiling method for the analysis of polar analytes in urine using HILIC-MS and ion mobility enabled HILIC-MS

开发一种使用 HILIC-MS 和离子淌度 HILIC-MS 分析尿液中极性分析物的快速分析方法

阅读:8
作者:Adam M King, Lauren G Mullin, Ian D Wilson, Muireann Coen, Paul D Rainville, Robert S Plumb, Lee A Gethings, Garth Maker, Robert Trengove

Conclusion

A rapid metabolic profiling method for the analysis of polar metabolites has been developed. The method has the advantages of speed and both reducing sample and solvent consumption compared to conventional profiling methods. The addition of IMS added an additional dimension for feature detection and the identification of metabolites.

Methods

A rapid untargeted HILIC LC-MS/MS profiling method for the analysis of small polar molecules has been developed. The 3.3 min separation used a Waters BEH amide (1 mm ID) analytical column on a Waters Synapt G2-Si Q-Tof enabled with ion mobility spectrometry (IMS). The methodology, was applied to the metabolic profiling of a series of rodent urine samples from vehicle-treated control rats and animals administered tienilic acid. The same separation was subsequently linked to IMS and MS to evaluate the benefits that IMS might provide for metabolome characterisation.

Results

The rapid HILIC-MS method was successfully applied to rapid analysis of rat urine and found, based on the data generated from the data acquired for the pooled quality control samples analysed at regular intervals throughout the analysis, to be robust. Peak area and retention times for the compounds detected in these samples showed good reproducibility across the batch. When used to profile the urine samples obtained from vehicle-dosed control and those administered tienilic acid the HILIC-MS method detected 3007 mass/retention time features. Analysis of the same samples using HILIC-IMS-MS enabled the detection of 6711 features. Provisional metabolite identification for a number of compounds was performed using the high collision energy MS/MS information compared against the Metlin MS/MS database and, in addition, both calculated and measured CCS values from an experimentally derived CCS database.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。