CE-MALDI interface based on inkjet technology

基于喷墨技术的 CE-MALDI 接口

阅读:9
作者:Michael W Vannatta, Colin D Whitmore, Norman J Dovichi

Abstract

An ink jet printer valve and a nozzle were used to deliver matrix and sample from an electrophoresis capillary onto a MALDI plate. The system was evaluated by the separation of a set of standard peptides. That separation generated up to 40 000 theoretical plates in less than 3 min. Detection limits were 500 amol for an ABI TOF-TOF instrument and 2 fmol for an ABI Q-TOF instrument. Over 70% coverage was obtained for the tryptic digest of alpha-lactalbumin in less than 2.5 min.