Ravulizumab: Characterization and quantitation of a new C5 inhibitor using isotype specific affinity purification and high-resolution mass spectrometry

Ravulizumab:使用同种型特异性亲和力纯化和高分辨率质谱法对新型 C5 抑制剂进行表征和定量

阅读:8
作者:Paula M Ladwig, Maria A V Willrich

Conclusion

The TOF-MS is a versatile development platform that can be used to characterize and verify the molecular weight differences between the ECUL and RAVUL heavy chains. Routine laboratory testing for RAVUL was viable using an orbitrap-MS to quantitate using the mass of the intact light chain. These two platforms, combined, provide incomparable value in development of LDTs for the clinical laboratory.

Methods

A time-of-flight mass spectrometer (TOF-MS) was used to characterize and differentiate the molecular weight differences between RAVUL and ECUL by both digest and reduction experiments. In parallel, an LDT for RAVUL was validated and implemented utilizing IgG4 enrichment with light chain detection and quantitation on a high throughput orbitrap MS platform.

Results

The TOF-MS platform allowed for the mass difference between RAVUL and ECUL to be verified along with providing a proof of concept for a new intact protein quantitation software. An LDT on an orbitrap MS was validated and implemented using intact light chain quantitation, with the limitation that it cannot differentiate between ECUL and RAVUL. The LDT has an analytical measuring range from 5 to 600 mcg/mL, inter-assay imprecision of ≤13% CV (n = 13) and accuracy with <4% error from expected values (n = 20).