Optofluidic flow meter for sub-nanoliter per minute flow measurements

用于每分钟亚纳升流量测量的光流体流量计

阅读:9
作者:Jalal Sadeghi, Paul N Patrone, Anthony J Kearsley, Gregory A Cooksey

Aim

We focus on improving an optofluidic system for measuring and calibrating microflows to the sub-nanoliter per minute range. Approach: Measurements rely on an optofluidic system that delivers excitation light and records fluorescence in a precise interrogation region of a microfluidic channel. Exploiting a scaling relationship between the flow rate and fluorescence emission after photobleaching, the system enables real-time determination of flow rates.

Conclusions

We demonstrate new capabilities in sub-nanoliter per minute flow control and measurement that are generalizable to cutting-edge light-material interaction and molecular diffusion for chemical and biomedical industries.

Results

Here, we demonstrate improved calibration of a flow controller to 1% uncertainty. Further, the resolution of the optofluidic flow meter improved to less than 1 nL / min with 5% uncertainty using a molecule with a 14-fold smaller diffusion coefficient than our previous report. Conclusions: We demonstrate new capabilities in sub-nanoliter per minute flow control and measurement that are generalizable to cutting-edge light-material interaction and molecular diffusion for chemical and biomedical industries.

Significance

Performance improvements in microfluidic systems depend on accurate measurement and fluid control on the micro- and nanoscales. New applications are continuously leading to lower volumetric flow rates. Aim: We focus on improving an optofluidic system for measuring and calibrating microflows to the sub-nanoliter per minute range. Approach: Measurements rely on an optofluidic system that delivers excitation light and records fluorescence in a precise interrogation region of a microfluidic channel. Exploiting a scaling relationship between the flow rate and fluorescence emission after photobleaching, the system enables real-time determination of flow rates. Results: Here, we demonstrate improved calibration of a flow controller to 1% uncertainty. Further, the resolution of the optofluidic flow meter improved to less than 1 nL / min with 5% uncertainty using a molecule with a 14-fold smaller diffusion coefficient than our previous report. Conclusions: We demonstrate new capabilities in sub-nanoliter per minute flow control and measurement that are generalizable to cutting-edge light-material interaction and molecular diffusion for chemical and biomedical industries.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。