Electronic field effect detection of SARS-CoV-2 N-protein before the onset of symptoms

症状出现前电子场效应检测 SARS-CoV-2 N 蛋白

阅读:8
作者:I Novodchuk, M Kayaharman, I Prassas, A Soosaipillai, R Karimi, I A Goldthorpe, E Abdel-Rahman, J Sanderson, E P Diamandis, M Bajcsy, M Yavuz

Abstract

As part of the efforts to contain the pandemic, researchers around the world have raced to develop testing platforms to detect the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), which causes the Coronavirus disease 2019 (COVID-19). Within the different detection platforms studied, the field effect transistor (FET) is a promising device due to its high sensitivity and fast detection capabilities. In this work, a graphene-based FET which uses a boron and nitrogen co-doped graphene oxide gel (BN-GO gel) transducer functionalized with nucleoprotein antibodies, has been investigated for the detection of SARS-CoV-2 nucleocapsid (N)-protein in buffer. This biosensor was able to detect the viral protein in less than 4 min, with a limit of detection (LOD) as low as 10 ag/mL and a wide linear detection range stretching over 11 orders of magnitude from 10 ag/mL-1 μg/mL. This represents the lowest LOD and widest detection range of any COVID-19 sensor and thus can potentially enable the detection of infected individuals before they become contagious. In addition to its potential use in the COVID-19 pandemic, our device serves as a proof-of-concept of the ability of functionalized BN-GO gel FETs to be used for ultrasensitive yet robust biosensors.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。