Technical Note: A deuterated 13 C-urea reference for clinical multiparametric MRI prostate cancer studies including hyperpolarized pyruvate

技术说明:氘代 13 C-尿素参考用于临床多参数 MRI 前列腺癌研究,包括高极化丙酮酸

阅读:10
作者:Collin J Harlan, Zhan Xu, Keith A Michel, Christopher M Walker, Sanjaya D Lokugama, Gary V Martinez, Mark D Pagel, James A Bankson

Conclusion

A deuterated reference reduced hyperintense ringing artifacts in 1 H images by reducing the 1 H signal produced from the 13 C-urea in the reference. The deuterated reference can be used to improve anatomical image quality in future clinical 1 H and hyperpolarized [1-13 C]-pyruvate MRI prostate imaging studies.

Methods

Due to a high amount of Gd doping for the purpose of reducing the spin-lattice relaxation time (T1 ) of urea, the 1 H signal produced by a reference of 13 C-urea in normal water was rapidly relaxed, resulting in severe artifacts in heavily T1 -weighted images. Hyperintense ringing artifacts in 1 H images were mitigated by reducing the 1 H concentration in a 13 C-urea reference via deuteration and lyophilization. Several references were fabricated and their SNR was compared using 1 H and 13 C imaging sequences on a 3T MRI scanner. Finally, 1 H prostate phantom imaging was conducted to compare image quality and 1 H signal intensity of normal and deuterated urea references.

Purpose

Metabolic magnetic resonance imaging (MRI) using hyperpolarized [1-13 C]-pyruvate offers unprecedented new insight into disease and response to therapy. 13 C-enriched reference standards are required to enable fast and accurate calibration for 13 C studies, but care must be taken to ensure that the reference is compatible with both 13 C and 1 H acquisitions. The goal of this study was to optimize the composition of a 13 C-urea reference for a dual-tuned 13 C/1 H endorectal coil and minimize imaging artifacts in metabolic and multiparametric MRI studies involving hyperpolarized [1-13 C]-pyruvate.

Results

The deuterated 13 C-urea reference provides strong 13 C signal for calibration and an attenuated 1 H signal that does not interfere with heavily T1 -weighted scans. Deuteration and lyophilization were fundamental to the reduction in 1 H signal and hyperintense ringing artifacts. There was a 25-fold reduction in signal intensity when comparing the nondeuterated reference to the deuterated reference, while the 13 C signal was unaffected.