Counter-Current Chromatographic Separation of Nucleic Acid Constituents with a Polar Volatile Organic-Aqueous Two-Phase Solvent Systems with ELSD Detection

采用极性挥发性有机-水双相溶剂系统进行逆流色谱分离核酸成分并进行 ELSD 检测

阅读:9
作者:Yoichi Shibusawa, Go Morikawa, Akio Yanagida, Yoichiro Ito

Abstract

Highly hydrophilic volatile organic/aqueous two-phase solvent systems containing an organic salt such as, acetonitrile/800 mM and 1200 mM ammonium acetate (1 : 1, v/v) were efficiently utilized for high-speed counter-current chromatography (HSCCC) to separate hydrophilic compounds. The retention of the upper and the lower stationary phases in the column of the cross-axis coil planet centrifuge (CCC instrument) was studied by changing the flow-rate of the mobile phase (1.0-3.0 ml/min). Using the acetonitrile/800 mM ammonium acetate two-phase solvent system, the stationary phase was retained at 46.3% relative to the total column capacity of 65 ml by the reversed-phase elution mode at a flow-rate of 1.0 ml/min. The best retention of the stationary upper phase of 51.5% was obtained by the solvent system of the acetonitrile/1200 mM ammonium acetate at the above flow-rate. With the acetonitrile/800 mM ammonium acetate system the base line separation of adenine and adenosine monophosphate (AMP) detected by evaporative light scattering detector (ELSD) and UV was achieved with lower phase mobile at a flow-rate of 1.0 ml/min within 70 min.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。