Background

Failure in the reprocessing of thermolabile flexible endoscopes has been reported as one of the most important threats to patient health. Method: A case report and observational study was conducted, from August 2014 to December 2019, in the Digestive Endoscopy Unit of a University Hospital in Italy, where two cases of Klebsiella pneumoniae carbapenemase (KPC)-producing Klebsiella pneumoniae infections in patients undergoing endoscopic retrograde cholangio-pancreatography were observed. Following the risk/safety management practices, an epidemiological investigation was started, duodenoscopes were removed from use and the reprocessing practices reviewed. Moreover, microbiological surveillance of endoscopes was carried out according to the CDC guidelines.

Conclusion

Managing the risk associated with the reprocessing of digestive endoscopes, through risk assessment at every stage of the process, is important for the prevention of infections associated with the use of these device.

Results

In the first phase of sampling, 10/10 (100%) endoscopes were found to be non-compliant, of which 7 showed results for high-concern organisms (HCOs), such as KPC-K. pneumoniae, P. aeruginosa and E. coli. After implementing corrective actions, 12 out of 17 endoscopes were found to be non-compliant (70.5%), of which 8 showed results for HCOs, such as KPC-K. oxytoca and P. aeruginosa. During the last year of regular microbiological surveillance, only 23% of endoscopes (35/152) were found to be non-compliant, of which 7 showed results for HCOs, such as NDM-K. pneumoniae, P. aeruginosa and A. baumannii. The crucial issues were related to samples collected from the internal channels of duodenoscopes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。