Determination of Clomipramine using eco-friendly solid-contact ionophore-doped potentiometric sensor

使用环保型固体接触离子载体掺杂电位传感器测定氯米帕明

阅读:11
作者:Adel M Michael, Amr M Mahmoud, Nesma M Fahmy

Conclusion

This work was concerned with developing a green analytical method for the determination of Clomipramine. The proposed SC-ISE was mixed with graphene nanocomposite transducer interlayer. The graphene layer succeeded in preventing the formation of an aqueous layer so resulted in a stable, reproducible standard potential besides the rapid response time.

Methods

Disposable screen-printed carbon electrode was utilized as a substrate to fabricate the proposed sensors. The sensors were optimized to determine Clomipramine using calix[4]arene as an ionophore into PVC polymeric membrane to enhance selectivity towards the target analyte. The solid-contact sensor potential stability was improved by the incorporation of graphene nanoparticles transducer layer.

Results

The sensors were assessed as per the IUPAC recommendations. The linearity range was 1 × 10- 2 to 1 × 10- 5.3 M. The sensors were successfully applied to determine CLM in the pharmaceutical formulation. Furthermore, the ion selective electrodes were applied for Clompiramine assay in spiked plasma for the purpose of Point-of-Care testing to be a diagnostic tool for therapeutic monitoring of the cited central nervous system agent. The findings were statistically compared to the reported method showing no statistically significant difference.