A simple method to overcome the inhibitory effect of heparin on DNA amplification

克服肝素对 DNA 扩增抑制作用的简单方法

阅读:16
作者:P G Vaughan-Shaw, M Walker, L Ooi, N Gilbert, S M Farrington, M G Dunlop

Background

Genetic material from large patient cohorts is increasingly central to translational genetic research. However, patient blood samples are a finite resource and their supply and storage are often dictated by clinical and not research protocols. Our experience supports difficulty in amplifying DNA from blood stored in herparin; a scenario that other researchers may have or will encounter. This technical note describes a number of simple steps that enable successful PCR amplification.

Conclusion

From our experience a standard phenol/chloroform purification and using the Invitrogen™ Platinum® Taq has enabled the amplification of whole blood samples taken into lithium heparin and stored frozen for up to a month. This simple method may enable investigators to utilise blood taken in lithium heparin for DNA extraction and amplification.

Methods

DNA was extracted using the Illustra Nucleon Genomic DNA Extraction Kit. PCR amplification was attempted using a number of commercially available PCR mastermixes.

Results

PCR DNA amplification failed using ReddyMix™ PCR Master Mix, Thermo-Start® (Thermo Scientific Inc. US) and ZymoTaq™ (Zymo research, US) PCR mastermixes, as demonstrated absence of products on gel electrophoresis. However, using the Invitrogen™ (Thermo Scientific Inc., US) Platinum® Taq DNA Polymerase, PCR products were identified on a 1% agarose gel for all samples. PCR products were cleaned with ExoSAP-IT® (Affymetrix Inc., US) and a sequencing reaction undertaken using a standard Big Dye protocol. Subsequent genotyping was successful for all samples for alleles at the CDH1 locus.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。