IQVision: An Image-Based Evaluation Tool for Quantitative Lateral Flow Immunoassay Kits

IQVision:基于图像的定量横向流动免疫测定试剂盒评估工具

阅读:8
作者:Lalitha Pratyusha Bheemavarapu, Malay Ilesh Shah, Jayaraj Joseph, Mohanasankar Sivaprakasam

Abstract

The development of quantitative lateral flow immunoassay test strips involves a lot of research from kit manufacturers' standpoint. Kit providers need to evaluate multiple parameters, including the location of test regions, sample flow speed, required sample volumes, reaction stability time, etc. A practical visualization tool assisting manufacturers in this process is very much required for the design of more sensitive and reliable quantitative LFIA test strips. In this paper, we present an image-based quantitative evaluation tool determining the practical functionality of fluorescence-labelled LFIA test cartridges. Image processing-based algorithms developed and presented in this paper provide a practical analysis of sample flow rates, reaction stability times of samples under test, and detect any abnormalities in test strips. Evaluation of the algorithm is done with Glycated Hemoglobin (HbA1C) and Vitamin D test cartridges. Practical sample flow progress for HbA1C test cartridges is demonstrated. The reaction stability time of HbA1C test samples is measured to be 12 min, while that of Vitamin D test samples is 24 min. Experimental evaluation of the abnormality detection algorithm is carried out, and sample flow abnormalities are detected with 100% accuracy while membrane irregularities are detected with 96% accuracy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。