Parallel, minimally-invasive implantation of ultra-flexible neural electrode arrays

超柔性神经电极阵列的平行、微创植入

阅读:11
作者:Zhengtuo Zhao, Xue Li, Fei He, Xiaoling Wei, Shengqing Lin, Chong Xie

Objective

Implanted microelectrodes provide a unique means to directly interface with the nervous system but have been limited by the lack of stable functionality. There is growing evidence suggesting that substantially reducing the mechanical rigidity of neural electrodes promotes tissue compatibility and improves their recording stability in both the short- and long-term. However, the miniaturized dimensions and ultraflexibility desired for mitigating tissue responses preclude the probe's self-supported penetration into the brain tissue. Approach: Here we demonstrate the high-throughput implantation of multi-shank ultraflexible neural electrode arrays with surgical footprints as small as 200 µm2 in a mouse model. This is achieved by using arrays of tungsten microwires as shuttle devices, and bio-dissolvable adhesive polyethylene glycol (PEG) to temporarily attach a shank onto each microwire. Main

Results

We show the ability to simultaneously deliver electrode arrays in designed patterns, to adjust the implantation locations of the shanks by need, to target different brain structures, and to control the surgical injury by reducing the microwire diameters to cellular scale. Significance: These results provide a facile implantation method to apply ultraflexible neural probes in scalable neural recording.

Significance

These results provide a facile implantation method to apply ultraflexible neural probes in scalable neural recording.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。